Sign In | Join Free | My udpf.com |
|
Categories | Stationary Meter Test Bench |
---|---|
Brand Name: | GENY |
Model Number: | YC1893D-ICT |
Certification: | ISO, CE |
Place of Origin: | China |
MOQ: | 1 |
Price: | negotiate rates |
Payment Terms: | T/T, L/C, D/P, T/T |
Supply Ability: | 5 sets / 1 months |
Delivery Time: | 6~8weeks |
Packaging Details: | Plywood cases |
Accuracy: | 0.05 |
Number Of Meter Positions: | 6/12/24/48 |
Structure: | All-in-one |
Power Source: | YCS-103-250 YCS-103-600 YCS-103-1200 YCS-103-2500 |
Test voltage output (Phase-Neutral): | 3x(24V~288V)(or customized) |
Stability: | Better than 0.005%/h (integration time 150 s) |
Material: | Aluminum Alloy |
Company Info. |
Guangzhou GENY Electric Co., Ltd |
Verified Supplier |
View Contact Details |
Product List |
>
Technical spec. | YC1893D-ICT-6 | YC1893D-ICT-12 | YC1893D-ICT-24 | YC1893D-ICT-48 |
Number of meter positions equipped ICT | 6 | 12 | 24 | 48 |
Accuracy class of ICT | 0.05 / 0.01 | |||
Accuracy | 0.05 | |||
Power Source | YCS-103-250 | YCS-103-600 | YCS-103-1200 | YCS-103-2500 |
Standard Meter | SZ03A-K6 / SZ03A-K8 (customized) | |||
Scanning Head | YCG-S Series | |||
Voltage | ||||
Test voltage output(Phase-Neutral) | 3x(24V~288V)(or customized) | |||
Power of the voltage output | Max. 3x100VA | Max. 3x250VA | Max. 3x500VA | Max. 3x1000VA |
Resolution | Better than 0.01% of full scale value of range | |||
Setting accuracy | Better than 0.05% at the final range value | |||
Stability | Better than 0.005% /h (integration time 150s) | |||
Load regulation from 0~max. load | Better than 0.01% | |||
Distortion factor | Better than 0.3% for linear resistance load | |||
Harmonic | 2nd~21st free programmable | |||
Current | ||||
Test current output | 3x(1mA~100A) | |||
Power of the current output | Max. 3x150VA | Max. 3x350VA | Max. 3x750VA | Max. 3x1500VA |
Resolution | Better than 0.01% of full scale value of range | |||
Setting accuracy | Better than 0.05% at the final range value | |||
Stability | Better than 0.005% /h (integration time 150s) | |||
Load regulation from 0~max. load | Better than 0.01% | |||
Distortion factor | Better than 0.3% for linear resistance load | |||
Harmonic | 2nd~21st free programmable | |||
Phase angle | ||||
Range | 0~360⁰ | |||
Resolution | 0.01⁰ | |||
Setting accuracy | 0.1⁰ | |||
Frequency | ||||
Range | 45Hz~65Hz | |||
Resolution | 0.01Hz | |||
Error display | ||||
Type of error display | Red LED | |||
Resolution of error displays | 8 digitals | |||
Other | ||||
Voltage supply | 3x220/380V±10%(or customized) 50/60Hz ±2Hz | |||
Ambient temperature | -10°C~+40°C | |||
Relative humidity | 35%~85% |
1. High quality product
2. Over 18 months long warranty
3. High accuracy class: 0.05
*** highest accuracy can reaches to 0.01% under 3 meter
positions***
4. 3~48 Three Phase Electric Energy Meter Test bench
5. Extremely high power stability
6. Timing error test
7. Influence test of voltage, frequency, and harmonic
8. Installing RS232, Ethernet communication port in each meter position.
9. Tilt impact test (optional)
![]() |